Explosion proof Inline Viscosity And Density Meter

Explosion proof Inline Viscosity And Density Meter

Designed for hazardous areas, this product complies with ATEX II1G and IECEx Zone 0 standards and measures the density and viscosity of liquids in microelectromechanical systems (MEMS). Within a MEMS system, the liquid is directed to an omega-shaped microchannel, the so-called omega chip. This extremely thin silicon tube, which is only as thick as a human hair, is measured by vibration. The density of a liquid can be estimated from the natural frequency of vibration: the denser the medium, the lower the vibration frequency.
Thanks to the MEMS system, the sensor measures only 37 x 83 x 18.2 mm³ and can be installed even in confined spaces. Its high-precision measurement results are available in real-time, supporting continuous measurement in the process. To compensate for the fact that the density of the liquid changes with temperature, a platinum resistor is integrated into the sensor to detect the temperature of the liquid.

產品特色

1. Modbus RTU communication
2. RS-485 point-to-point connection
3. Based on the Modbus serial communication specification
4. In-line measurement
5. Direct integration: Can be directly embedded in running processes
6. Instant data output: Measured values are output instantaneously
7. Fast media changeover: Enables quick change of measuring medium
8. Density measurement
9. Density value
10. Maximum Measurement Deviation:
11. ±0.2 kg/m³ or ± [0.0075 x abs (T-25 °C)] kg/m³ (if deviation > 0.2 kg/m³)
12. Repeatability: ±0.1 kg/m³
13. Viscosity measurement
14. Viscosity values: dynamic viscosity and kinematic viscosity
15. Maximum measurement error: ± [0.2 mPa·s + 5% of measured value]
16. Repeatability: ±0.1 mPa·s
17. Concentration measurement is optional
18. Various types of sugars in water
19. Invert sugar in water
20. High fructose corn syrup
21. Methanol in water
22. Ethanol in water
23. Salt in water
24. Minerals in water
25. Hydrogen peroxide in water

檔案下載

規格

Permissible density measurement range :
0… 1600 kg/m³
Permissible viscosity range :
0.3… 50 mPa·s
Permissible medium pressure:
0… 20 bar (absolute)
Burst pressure: 80 bar (absolute)
Allowable particle size :
Up to 30 μm
Allowable flow range :
0… 10 L/h (water)
Vibration:
Due to the high operating frequency of the microchannel, vibration (<20 kHz) has no effect on the measurement accuracy.
Import and export section :
The inlet and outlet segments have no effect on the measurement accuracy.

  • 型號:VLO-M2_ex